  Alnor AirGard® 385 Fume Hood Monitor

Features

The AirGard 385 continuously monitors fume hood face velocity by measuring the air velocity of clean air from the room passing through the air inlet on the front of the monitor. The unit then reports this information visually through a digital display and colored lights, as well as with a continuous audible alarm when the face velocity is in an alarm condition. This information can be sent to a remote location by means of a dry contact relay output and 0-4 volt analog output. The monitor mounts flush to the fume hood surface and fits into a single electrical receptacle box knock-out.

Specifications

Digital Display 3digit, 7-segment liquid crystal display.
Display Units Feet per minute (fpm) or meters per second (m/s). User selectable through menu.
Display Range 70 to 250 fpm (.35 to 1.25 m/s). Over range measurements shown as Hi. Under range measurements shown as Lo.
Resolution 1 fpm or .01 m/s.
Accuracy ±10% or 10 fpm, whichever is greater.
Alarm Range 70 to 250 fpm (.35 to 1.25 m/s).
Alarm Delays User configurable through menu from 0 to 255 seconds.
Audible Alarm Indication 85dB @ 4 inches (10 cm).
Visual Alarm Indication Red LED.
Visual Warning Indications Yellow LEDs for low flow and high flow warnings zones. Warning offsets are user configurable through menu.
Relay Output Nominal switching capacity 1A @ 30V DC, .5A @ 125V AC, form C relay.
High Flow Alarm Disable Audible, visual and relay disable.
Horn Silence Temporary & permanent.
Night Set back Unit can be disabled remotely.
User Alarm Input Intermittent audible and LCD status indicator indicate external alarm.
Analog Output 0 to 4V DC.
Calibration Two point field calibration required.
Instrument Dimensions
Front Faceplate 5.25L x 3.0W x .5D inch (13.3L x 7.6W x 1.3D cm).
Rear Enclosure 3.0L x 2.0W x .5D inch (7.6L x 5.0W x 1.3D cm).
Mounting Flush. (3.0L x 2.0W inch cut out required on hood surface).
Monitor fits into a single electrical receptacle box knock-out.
Operating Conditions 55 to 86°F (13 to 30°C), 5% to 95% RH, non-condensing.
Storage Temperature -40 to 150°F (-40 to 65°C), 5% to 95% RH, non-condensing.
Power Requirement 9 to 24V AC/DC. (Wall plug in power supply supplied for domestic units).
The calibration and configurable parameters are stored to nonvolatile memory of the instrument and are not lost when the unit loses power.

Model

Model 385
Flush mount. Digital flow indication (high and low alarm).
Includes mounting hardware, tubing, sidewall adapter,
wall plug-in power supply and owner's manual.
670-385-000

Please select from the drop down menu below
Base price : US$495.00
Model :
Unit Price:
Quantity:

© 2006 Omni Controls, Inc.. All Rights Reserved.

Email: Sales@Omnicontrols.com
We guarantee a 24 hour response on all inquiries made via email.
For quicker response please call Omni Controls
at any of the above numbers